Incontinence Stimulator

$124.00
Qty:

The EM-2400 incontinence stimulator is a battery operated pulse generator that stimulates the muscles of the pelvic floor by means of a vaginal or rectal electrode. The unit has a single controllable channel which the electrode is connected to.

The stimulator has 5 preset programmes. The programme type and the intensity of stimulation are controlled by push buttons. 
 
 

 

MECHANISM

TECHICAL DESCRIPTION

01

Channel
Single channel

02

Pulse Amplitude
Adjustable, 0-80mA, Max output 80mA (peak to peak) into 500ohm load.

03

Output Voltage
Adjustable, 0-40V, Max output 40V (peak to peak) into 500ohm load.

04

Wave Form
Asymmetrical Rectangular Biphasic Pulse.

05

Power Supply
Two AA Batteries

06

Size
9.2cm(L) x 6.2cm(W) x 2.85cm(H)

07

Weight
145 grams (2 batteries included)

08

Pulse Rate
Pre-setting, 1~100 Hz

09

Pulse Width
Pre-setting, 150 µs, 200 µs or 250 µs

10

Programme
A, B, C, D, E, five pre-set programme of options.

11

Programme 
A ,B, C, E
The programme details of A, B, C, E are as follows. Parameters are preset at fixed values. The 2 seconds ramp time includes 1 second ramp up, and 1 second ramp down. Each programme is repeated in a constant cyclic pattern.
Programme
Rate
Width
Contraction
Relaxation
Ramp
Cycle
 
(Hz)
(µs)
(Sec.)
(Sec.)
(Sec.)
(Sec.)
A
35
250
4
8
2
14
B
40
200
6
15
2
23
C
10
250
5
3
2
10
E
20
250
5
5
2
12

12

Programme D
Programme D is composed of 12 sections with varied pulse rates. Each section contains 30 contractions. The 2 seconds ramp time includes 1 second ramp up, and 1 second ramp down. The complete programme lasts 60 minutes with 360 contractions. It is repeated in a constant cyclic pattern. The details of the programme are given below.
Section
Rate
Width
Contraction
Relaxation
Ramp
Cycle
 
(Hz)
(µs)
(Sec.)
(Sec.)
(Sec.)
(Sec.)
1
7
150
4
4
2
10
2
8
150
4
4
2
10
3
9
150
4
4
2
10
4
10
150
4
4
2
10
5
11
150
4
4
2
10
6
12
150
4
4
2
10
7
13
150
4
4
2
10
8
14
150
4
4
2
10
9
15
150
4
4
2
10
10
16
150
4
4
2
10
11
12
150
4
4
2
10
12
7
150
4
4
2
10

13

Timer
Fixed at 30 minutes, treatment time countdown automatically. The output current will be shut off when the time is expired followed by  a beep sound.

14

Operating Condition
Temperature:0° C ~ 40° C 
Relative Humidity: 30%~75%
Atmosphere Pressure: 700Hpa~1060Hpa

15

Remark
There may be up to a +/-10% tolerance of all parameters and +/-20% tolerance of amplitude & voltage.